| Model |
HCP
108 |
HCP
153 |
HCP
246 |
| Stainless steel interior
(deep-drawn electropolished) |
Volume |
approx. l |
108 |
153 |
246 |
| With/Height/Depth (less
25 mm for air duct in the middle of the back wall) |
mm |
560/480/400 |
480/640/500 |
640/640/600 |
| Provision for sliding stainless steel
shelves resp. wire grid shelves |
number |
4 |
6 |
6 |
| Stainless-steel
exterior (rear zinc-plated steel) |
Width |
mm |
710 |
630 |
790 |
| Height |
mm |
778 |
938 |
938 |
| Depth
(without door handle, depth of handle 38 mm) |
mm |
550 |
650 |
750 |
| Fully insulated,
heated stainless steel door |
|
x |
x |
x |
| Extra internal glass
door |
|
x |
x |
x |
| Ventilation |
Uniform atmosphere and temperature
distribution through enclosed non-turbulent ventilation system,
fully covered by the sterilisation process |
|
x |
x |
x |
| Humidity |
Capacitive
humidity sensor (sterilisable) |
|
x |
x |
x |
| Active
microprocessor control for humidifying and dehumidifying (20-95 % rh),
incl. digital indication and auto-diagnostic system ensures even
more rapid reaching of set humidity and very short recovery times
while avoiding condensate formation; humidity supply with distilled
water (from an external tank) by a self-priming pump; integral
bacteria block by generating hotsteam, dehumidifying via sterile
filter |
|
x |
x |
x |
| Temperature |
Electronic microprocessor temperature
controller with PT 100 and auto-diagnostic system |
|
x |
x |
x |
| Temperatures sensor PT100
Class A in 4-wire circuit for uninterrupted operation on failure of
one PT100 with warning indication |
|
double |
double |
double |
Temperature range with
humidity control:
Temperature range without humidity control:
(during sterilisation the temperature is fixed at 160 °C - set
value)) |
°C
°C |
from 20 up to
90
from 20 up to 160 |
from 20 up to 90
from 20 up to 160
|
from 20 up to
90
from 20 up to 160 |
Temperature
fluctuations with time
Temperature variation in chamber at 90 °C |
°C
°C |
£
± 0,1
£
± 0,5 |
£
± 0,1
£
± 0,5
|
£
± 0,1
£
± 0,5 |
| Sterilisation |
STERICard
for automatic chamber sterilisation cycle 4 h at 160 °C (not for
sterilising the load) |
|
x |
x |
x |
| Humidity |
Capacitive
humidity sensor (sterilisable) |
|
x |
x |
x |
| Active
microprocessor control for humidifying and dehumidifying (20-95 % rh),
incl. digital indication and auto-diagnostic system ensures even
more rapid reaching of set humidity and very short recovery times
while avoiding condensate formation; humidity supply with distilled
water (from an external tank) by a self-priming pump; integral
bacteria block by generating hotsteam, dehumidifying via sterile
filter |
|
x |
x |
x |
| Monitor |
Microprocessor
temperature monitor acting as overtemperature protection, with PT
100 incorporating fault diagnostics with visual and audible alarm |
|
x |
x |
x |
| Digital
over- and under-temperature monitor |
|
x |
x |
x |
| Relay for
reliable heating cut-off in case of fault |
|
x |
x |
x |
| Mechanical
temperature limiter (TB) |
|
x |
x |
x |
| Audible
alarm: Over- and undertemperature; over-/underhumidity: open door
and empty water tank |
|
x |
x |
x |
| Setpoint
linked gliding automatic safety function |
|
x |
x |
x |
| Timer functions |
Timer with residual running
time: max. 40 ramps (each 1 min. up to 999 h) programmable through
controller or MEMoryCard XL; programming via PC and free-of-charge
software: unlimited number of ramps |
|
x |
x |
x |
| Real-time/weekly programmer
with group function |
|
x |
x |
x |
| Documentation |
Internal
documentation memory 1024 kB as ring memory for all setpoints,
actual values, errors, settings with real-time and date; capacity
approx. 3 months at 1 min. intervals |
|
x |
x |
x |
| Parallel
printer interface for printing documentation data, suitable for all
PCL3-compatible ink jet printers |
|
x |
x |
x |
| "Celsius
2007" software for control and documentation of temperature
and relative humidity |
|
x |
x |
x |
| Setup |
Calibration (no separate PC
required): Temperature: 3-point calibration on controller; Humidity: 2-points calibration at 20 % and 90 % |
|
x |
x |
x |
| Setting of language for
dialogue resp. display |
|
x |
x |
x |
|
|
|
|
|
|
| Further
data |
Electrical supply |
V/Hz |
230 /
50/60 |
230 /
50/60 |
230 /
50/60 |
| Electrical load |
approx. W |
1000 |
1500 |
2000 |
| Net weight |
approx. kg |
70 |
82 |
110 |
| Standard accessories |
Stainless steel sliding shelves |
number |
2 |
2 |
2 |
| Shelf or wire grid shelf
width |
approx. mm |
560 |
480 |
640 |
| Shelf depth |
approx. mm |
300 |
400 |
500 |
| Works calibration
certificate (test point chamber centre at 60 °C) |
|
x |
x |
x |
|
|
x = Standard model, basic equipment |
|
|
|
|