Memmert-Humidity Chamber HCP

Model HCP 108 HCP 153 HCP 246
Stainless steel interior (deep-drawn electropolished) Volume approx. l 108 153 246
With/Height/Depth (less 25 mm for air duct in the middle of the back wall) mm 560/480/400 480/640/500 640/640/600
Provision for sliding stainless steel shelves resp. wire grid shelves number 4 6 6
Stainless-steel exterior (rear zinc-plated steel) Width mm 710 630 790
Height mm 778 938 938
Depth (without door handle, depth of handle 38 mm) mm 550 650 750
Fully insulated, heated stainless steel door   x x x
Extra internal glass door   x x x
Ventilation Uniform atmosphere and temperature distribution through enclosed non-turbulent ventilation system, fully covered by the sterilisation process   x x x
Humidity Capacitive humidity sensor (sterilisable) x x x
Active microprocessor control for humidifying and dehumidifying (20-95 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times while avoiding condensate formation; humidity supply with distilled water (from an external tank) by a self-priming pump; integral bacteria block by generating hotsteam, dehumidifying via sterile filter   x x x
Temperature Electronic microprocessor temperature controller with PT 100 and auto-diagnostic system x x x
Temperatures sensor PT100 Class A in 4-wire circuit for uninterrupted operation on failure of one PT100 with warning indication double double double
Temperature range with humidity control:
Temperature range without humidity control:
(during sterilisation the temperature is fixed at 160 °C - set value))
°C
°C
from 20 up to 90
from 20 up to 160

from 20 up to 90
from 20 up to 160

from 20 up to 90
from 20 up to 160
Temperature fluctuations with time 
Temperature variation in chamber at 90 °C
°C
°C
£ ± 0,1
£ ± 0,5
£ ± 0,1
£ ± 0,5

 

£ ± 0,1
£ ± 0,5
Sterilisation STERICard for automatic chamber sterilisation cycle 4 h at 160 °C (not for sterilising the load)   x x x
Humidity Capacitive humidity sensor (sterilisable)   x x x
Active microprocessor control for humidifying and dehumidifying (20-95 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times while avoiding condensate formation; humidity supply with distilled water (from an external tank) by a self-priming pump; integral bacteria block by generating hotsteam, dehumidifying via sterile filter   x x x
Monitor Microprocessor temperature monitor acting as overtemperature protection, with PT 100 incorporating fault diagnostics with visual and audible alarm   x x x
Digital over- and under-temperature monitor   x x x
Relay for reliable heating cut-off in case of fault   x x x
Mechanical temperature limiter (TB)   x x x
Audible alarm: Over- and undertemperature; over-/underhumidity: open door and empty water tank   x x x
Setpoint linked gliding automatic safety function   x x x
Timer functions Timer with residual running time: max. 40 ramps (each 1 min. up to 999 h) programmable through controller or MEMoryCard XL; programming via PC and free-of-charge software: unlimited number of ramps x x x
Real-time/weekly programmer with group function x x x
Documentation Internal documentation memory 1024 kB as ring memory for all setpoints, actual values, errors, settings with real-time and date; capacity approx. 3 months at 1 min. intervals   x x x
Parallel printer interface for printing documentation data, suitable for all PCL3-compatible ink jet printers   x x x
"Celsius 2007" software for control and documentation of temperature and relative humidity   x x x
Setup Calibration (no separate PC required): Temperature: 3-point calibration on controller; Humidity: 2-points calibration at 20 % and 90 % x x x
Setting of language for dialogue resp. display x x x
Further data Electrical supply V/Hz 230 / 50/60 230 / 50/60 230 / 50/60
Electrical load approx. W 1000 1500 2000
Net weight approx. kg 70 82 110
Standard accessories Stainless steel sliding shelves number 2 2 2
Shelf or wire grid shelf width  approx. mm 560 480 640
Shelf depth approx. mm 300 400 500
Works calibration certificate (test point chamber centre at 60 °C)    x x x

 

x = Standard model, basic equipment
 
 