Hettich AG
Seestrasse 204a
CH-8806 Bäch
Tel.: +41 44 786 80 20
Fax: +41 44 786 80 21
e-Mail: info@hettich-ag.ch |
|
Automatic Cell Washing System
for Serological Testing
In
blood banks and transfusion centres, it is routine practice to test the
compatibility between donor and recipient prior to every blood
transfusion. The Rotolavit cell washing system is ideal for this viral
application. The washing process is carried out automatically. The
appropriate quantity of saline solution is filled equally into 12 or 24
tubes, depending on the rotor, and mixed with a previously pipetted drop
of blood. It is then centrifuged and the supernatant is decanted. The
washing cycle is repeated several times and the anti-human globulin
(Coombs) serum is then added manually, in order to assess possible
agglutination following the final centrifugation.
The washing cycle process is repeated automatically, allowing uü to 9
sequenced washing cycles to be programmed freely.
A washing cycle includes the following steps:
- Filling the saline solution
- Centrifuging at (usually) 3.500 min-1
- Decanting the supernatant
- Resuspending the sediment (between cycles only)
Digital display and control panel
Display
| PROG |
Entry of the program number, 4
programmable memories are available |
| CYC |
Entry of the number
of washing cycles (max. 9) |
| ml |
Entry of the saline solution quantity
per tube (max. 5 ml) |
| RPM |
Entry of the
rotational speed |
| t/min:s |
Entry of the washing cycle duration
(max. 9 min:59 sec) |
| An LED lights up: |
Should an imbalance
occur
As long as the rotor is turning
When the lid can be opened, once the program has been completet |
Keypad

| ARROW KEYS |
Leads the user through the menue
Increases the corresponding value
Decreases the corresponding value |
| SPIN |
Starts centrifugation
for agglutination |
| START WASH |
Starts the washing process |
| STOP |
Stops centrifugation
manually |
| SALINE |
Recalls the programmed quantity of
saline solution |
| CHECK |
Checks the quantity
of saline solution per washing cycle |
Areas of application
- Cross-matching
- Antibody search and differentiation
Technology
- Microprocessor control
- Frequency drive
- 4 programmable memories
- Max. RCF 1.438 (3.500 min-1)
User-friendliness
- Effortless cleaning:
simply run a few washing cycles with cleansing agent and then with
distilled water
- Easy rotor changing
- A visual and acoustic signal warns when the saline
solution container is empty
- One-handed lid closing
Safety
- Lid locking and holding
- Lid dropping protection
- Emergency lid lock release
- Metal housing
- Imbalance switch-off
- Motor cut-off at standstill
Specifications:
|
Rotolavit |
| Cell washing centrifuge, without rotor |
|
| Power supply |
220 - 240 V1 |
| Frequency Net/crated |
50 - 60 Hz |
| Consumption |
300 VA |
| Radio interference |
EN 55011 ISM Klasse B |
| Immunity |
EN 50082-1 |
|
|
| Max. capacity |
24 standard tubes (10x75 mm or 12x75
mm) |
| Max. RPM (speed) / RCF |
3500 min-1 / 1.438 |
| Radius (both rotors) |
105 mm |
| Max. centrifuging time |
max. 9 min:50 sec |
|
|
| Dimensions (HxWxD) in mm |
278 x 333 x 420 |
| Weight |
approx. 23 kg |
| Marking |