Automatic Cell Washing System 
for Serological Testing

 

In blood banks and transfusion centres, it is routine practice to test the compatibility between donor and recipient prior to every blood transfusion. The Rotolavit cell washing system is ideal for this viral application. The washing process is carried out automatically. The appropriate quantity of saline solution is filled equally into 12 or 24 tubes, depending on the rotor, and mixed with a previously pipetted drop of blood. It is then centrifuged and the supernatant is decanted. The washing cycle is repeated several times and the anti-human globulin (Coombs) serum is then added manually, in order to assess possible agglutination following the final centrifugation.

The washing cycle process is repeated automatically, allowing uü to 9 sequenced washing cycles to be programmed freely.

A washing cycle includes the following steps:

  1. Filling the saline solution
  2. Centrifuging at (usually) 3.500 min-1
  3. Decanting the supernatant
  4. Resuspending the sediment (between cycles only)

Digital display and control panel

Display

PROG Entry of the program number, 4 programmable memories are available
CYC Entry of the number of washing cycles (max. 9)
ml Entry of the saline solution quantity per tube (max. 5 ml)
RPM Entry of the rotational speed
t/min:s Entry of the washing cycle duration (max. 9 min:59 sec)
An LED lights up: Should an imbalance occur
As long as the rotor is turning
When the lid can be opened, once the program has been completet

Keypad

ARROW KEYS Leads the user through the menue
Increases the corresponding value
Decreases the corresponding value
SPIN Starts centrifugation for agglutination
START WASH Starts the washing process
STOP Stops centrifugation manually
SALINE Recalls the programmed quantity of saline solution
CHECK Checks the quantity of saline solution per washing cycle

Areas of application

  • Cross-matching
  • Antibody search and differentiation

Technology

  • Microprocessor control
  • Frequency drive
  • 4 programmable memories
  • Max. RCF 1.438 (3.500 min-1)

User-friendliness

  • Effortless cleaning:
    simply run a few washing cycles with cleansing agent and then with distilled water
  • Easy rotor changing
  • A visual and acoustic signal warns when the saline solution container is empty
  • One-handed lid closing

Safety

  • Lid locking and holding
  • Lid dropping protection
  • Emergency lid lock release
  • Metal housing
  • Imbalance switch-off
  • Motor cut-off at standstill

Specifications:

Rotolavit
Cell washing centrifuge, without rotor
Power supply 220 - 240 V1
Frequency Net/crated 50 - 60 Hz
Consumption 300 VA
Radio interference EN 55011 ISM Klasse B
Immunity EN 50082-1
Max. capacity 24 standard tubes (10x75 mm or 12x75 mm)
Max. RPM (speed) / RCF 3500 min-1 / 1.438
Radius (both rotors) 105 mm
Max. centrifuging time max. 9 min:50 sec
Dimensions (HxWxD) in mm 278 x 333 x 420
Weight approx. 23 kg
Marking CE

 